<Appendix 2 For SC (Conductivity)>
App.2-12
IM 12A01A02-01E 8th Edition : Oct. 01, 2015-00
The USP<645> limit value is a non-temperature-compensated conductivity. When a measuring
conductivity needs to be shown on the main or home display as a non-temperature-compensated
conductivity, set the temperature compensation on ā€œNoneā€. (Refer to the section 7.1.4.)
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
25 50 75 100
USP safety margin
(user programmable)
USP limit
Temperature (°C)
Conductivity (µS/cm)
Figure 3 Conductivity limit value and temperature directed by USP<645>
n Changing the settings
If any setting is accidentally changed, values to the right of the relevant arrow in Table 6 are all
initialized.
Table 6 Parameters that initialize other values
Measurement -> Output: Process
parameter ->
Linear: 0% value, 100% value
Table
Communication: HART: PV
Display setup: Individual display (Main display)
Trend Graph Screen -> Y-axis (low, high)
Communication: HART
Congure sensor:
Measuring unit ->
Display setup: Individual display (Main display): unit

Table of Contents

Other manuals for YOKOGAWA FLXA21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA FLXA21 and is the answer not in the manual?

YOKOGAWA FLXA21 Specifications

General IconGeneral
BrandYOKOGAWA
ModelFLXA21
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals