IM 12A01A02-01E 11/11
8th Edition
Addition of Type n
(page 2-5) addition of “2.3 Installing the cable glands”
WARNING
For Type n model (-DB, -DE),
- When the accompanying cable glands are used, external clamping is required additionally.
- When the optional conduit adaptors(/CB5, /CD5, /CF5) are used with accompanying cable glands,
clamping; e.g) sealing tting is required.
- If external clamping is not able to be prepared, Ex d, Ex e, or Ex n cable gland which provides sufcient
clamping shall be used instead of the accompanying cable gland.
(page 2-4) change of “2.2 Removing the wiring cover”
(page 2-10) change of “2.4.3 Wiring cover”
NOTE
Wiring covers are required intrinsic safety and Type n (Type is -C*, -DB, -DE). In the case of other types,
the cover is not attached to the FLXA202.
Customer Maintenance Parts List
Revised CMPL 12A01A02-01E (11th Edition) and CMPL 12A01A03-01EN (2nd Edition) are attached to this
IM supplement.