<2. WIRING AND INSTALLATION>
2-17
IM 12A01A02-01E 8th Edition : Oct. 01, 2015-00
NOTE
The special grommet is intended to be used to seal the multiple cables from the Yokogawa ow
ttings such as FF20.
The designated cables are WU20 sensor cables, which are approximately 5 mm (0.2”) in
diameter, and K1500FV liquid earth cables, which are approximately 2.5 mm (0.1”) in diameter.
For sensor systems using a single cable, like the FU20/FU24 and the PR10, PD20, PF20 and
PS20, the standard gland will accommodate the cable adequately. Single cables between
approximately 6 mm and 12 mm (0.24” and 0.47”) can be sealed properly with these glands and
the standard tule.
Remove
standard tule
Sensor
Grommet set
Figure 2.13 Grommet set use
PH
SC
ISC
DO
SENCOM
2