EasyManuals Logo

YOKOGAWA MY600 User Manual

YOKOGAWA MY600
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
52
IM MY600-01EN
4.
With the measurement switch locked or the remote
switchhelddown,press0ΩADJ.
lights,and0.00Ωisdisplayed.Thevaluewillbe
stored in memory and will not be cleared even when
the power is turned off.
5.
Toclearthevalue,press0ΩADJwiththeprobe
opened. When the value is cleared,
disappears.
When the resistance cannot be cancelled, “no” will be
displayed,and0Ωadjustmentwillnotbeexecutedeven
ifyoupress0ΩADJ.Ifyouwanttocanceltheresistance,
contact your nearest YOKOGAWA dealer to have the
instrumentadjusted.
Measurement Procedure
1.
SettherangeswitchtotheV/Ωposition.
2.
If the instrument is set to voltage measurement
mode, press SELECT to change to low-resistance
measurement mode.
3.
Connect the probe to the DUT, and press the
measurement switch or remote switch.
Measurement principle:
Resistance = Voltage/Current
RX = V/ I
Voltage: V
LINE(+)
EARTH(-)
Current: I
Resistance: RX
• Ifacurrentof200mAormorerunsduringthe
measurement, a buzzer sounds to indicate that power
is running. You can turn off this buzzer (see chapter 8,
“Backlight, LED Light, and Buzzer”).
• Themeasurementresultsoflow-resistance
measurement may be affected by the impedance of the
operating circuit connected in parallel.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MY600 and is the answer not in the manual?

YOKOGAWA MY600 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMY600
CategoryTest Equipment
LanguageEnglish

Related product manuals