EasyManua.ls Logo

YOKOGAWA MY600 - Low-Resistance Measurement (Continuity Check); 0Ω Adjustment Function

YOKOGAWA MY600
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51
IM MY600-01EN
7. Low-Resistance Measurement
(Continuity Check)
WARNING
Do not apply voltage in the low-resistance measurement
range. Check that the power is not running through the
circuit under test before making a measurement.
French
AVERTISSEMENT
Ne pas appliquer de tension à la plage de faible résistance.
Toujoursvérifierlecircuitoul’équipementsoustestest
désactivé de manière sûre avant de commencer une
mesure.
Adjustment Function
The0Ωadjustmentfunctioncancelstheresistancesofthe
probe and fuse to display only the resistance of the DUT.
Uptoabout3Ωofresistancecanbecancelled.
Setup Procedure
1.
SettherangeswitchtotheV/Ωposition.
2.
If the instrument is set to voltage measurement
mode, press SELECT to change to low-resistance
measurement mode.
3.
Short the red probe connected to the LINE terminal and
the black probe connected to the EARTH terminal.

Related product manuals