51
IM MY600-01EN
7. Low-Resistance Measurement
(Continuity Check)
WARNING
Do not apply voltage in the low-resistance measurement
range. Check that the power is not running through the
circuit under test before making a measurement.
French
AVERTISSEMENT
Ne pas appliquer de tension à la plage de faible résistance.
Toujoursvérifierlecircuitoul’équipementsoustestest
désactivé de manière sûre avant de commencer une
mesure.
Adjustment Function
The0Ωadjustmentfunctioncancelstheresistancesofthe
probe and fuse to display only the resistance of the DUT.
Uptoabout3Ωofresistancecanbecancelled.
Setup Procedure
1.
SettherangeswitchtotheV/Ωposition.
2.
If the instrument is set to voltage measurement
mode, press SELECT to change to low-resistance
measurement mode.
3.
Short the red probe connected to the LINE terminal and
the black probe connected to the EARTH terminal.