Page #80 background image
General Instruction Manual
Troubleshooting
Malfunction of operation
80 / 90
IM 01U10B00-00EN-R, 3rd edition, 2018-07-09
10 Troubleshooting
All error messages and error codes that may appear in operation are described in the
software instruction manual. Possible malfunctions that may occur during commissioning
are explained below and remedying them is explained. If you cannot remedy the malfunc-
tion using these explanations, contact the Yokogawa service center.
10.1 Malfunction of operation
Malfunction Possible causes Remedy
Display on
transmitter not
functioning
Power supply
disconnected
Ensure that the unit is connected to the
power supply, see [}63]
Settings cannot be
made via IR switches
Check cable connections between
display and main board and connect
properly, if necessary.
Settings cannot
be made via IR
switches
Incorrect settings in
write-protect menu
item
Switch off write-protect menu item via
digital communication or hardware
switch.
Field communica-
tor is not detected
HART or Modbus DD
not installed on field
communicator
Save HART or Modbus DD file on field
communicator.
Field communicator
not connected
Connect field communicator with
Rotamass Total Insight, see software
instruction manual.

Table of Contents

Other manuals for YOKOGAWA ROTAMASS Total Insight

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ROTAMASS Total Insight and is the answer not in the manual?

YOKOGAWA ROTAMASS Total Insight Specifications

General IconGeneral
BrandYOKOGAWA
ModelROTAMASS Total Insight
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals