1-5
!"#$"
Dextrose (D-Glucose)
YSI Membrane: 2365
Detection Range: 0-9 g/L at 25 µL sample size,
0-25 g/L at 10 µL sample size
Calibration Point: 2.50 g/L
Linearity Check Point: 9.0 g/L
Precision (CV,n=10): 2%
Linearity (0 to Cal Point): ±2%
Linearity (Cal to Range Max): ±5%
Typical Membrane Working Life: 21 days
Membrane O-ring Color: Red
Ethanol
YSI Membrane: 2786
Detection Range: 0-3.2 g/L
Calibration Point: 2.00 g/L
Linearity Check Point: 3.20 g/L
Precision (CV,n=10): 2%
Linearity (0 to Cal Point): ±2%
Linearity (Cal to Range Max): ±5%
Typical Membrane Working Life: 5 days
Membrane O-ring Color: Green
Potential Substrate Interference: Methanol
L-Glutamate*
YSI Membrane: 2754
Detection Range: 0-10 mmol/L
Calibration Point: 5.00 mmol/L
Linearity Check Point: 10.0 mmol/L
Precision (CV,n=10): 2%
Linearity (0 to Cal Point): ±2%
Linearity (Cal to Range Max): ±5%
Typical Membrane Working Life: 7 days
Membrane O-ring Color: Yellow
*User may configure instrument to measure as monosodium glutamate (MSG).
L-Lactate (L-Lactic Acid)
YSI Membrane: 2329
Detection Range: 0-2.67 g/L
Calibration Point: 0.50 g/L
Linearity Check Point: 2.67 g/L
Precision (CV,n=10): 2%
Linearity (0 to Cal Point): ±2%
Linearity (Cal to Range Max): ±5%
Typical Membrane Working Life: 14 days
Membrane O-ring Color: Gray