1-6
Lactose
YSI Membrane: 2702
Detection Range: 0-25 g/L
Calibration Point: 5.00 g/L
Linearity Check Point: 25.0 g/L
Precision (CV,n=10): 2%
Linearity (0 to Cal Point): ±2%
Linearity (Cal to Range Max): ±5%
Typical Membrane Working Life: 10 days
Membrane O-ring Color: White
Sucrose
YSI Membrane: 2703
Detection Range: 0-25 g/L
Calibration Point: 5.00 g/L
Linearity Check Point: 25.0 g/L
Precision (CV,n=10): 2%
Linearity (0 to Cal Point): ±2%
Linearity (Cal to Range Max): ±5%
Typical Membrane Working Life: 10 days
Membrane O-ring color: Blue
Choline
YSI Membrane: 2771
Detection Range: 0-450 mg/L
Calibration Point: 175 mg/L
Linearity Check Point: 450 mg/L
Precision (CV,n=10): 2%
Linearity (0 to Cal Point): ±2%
Linearity (Cal to Range Max): ±5%
Typical Membrane Working Life: 7 days
Membrane O-ring color: Orange
L-Glutamine*
YSI Membrane: 2735
Detection Range: 0-8 mmol/L
Calibration Point: 5.00 mmol/L
Linearity Check Point: 8.00 mmol/L
Precision (CV,n=5 ): 4%
Linearity (0 to Cal Point): ±4%
Linearity (Cal to Range Max): ±5%
Typical Membrane Working Life: 5 days
Membrane O-ring color: Magenta