5-12
Prompt replicate cycle? No
1-No 2-Yes
Press [ENTER] to confirm and return to Sampling protocol setup.
Sampling protocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
Press [3] for ID. If you wish to assign sample identification numbers to each sample, enter [2]
for Yes here. When you press [SAMPLE] in RUN Mode, the display will prompt you to enter a
sample ID number. You may elect not to assign a number for each sample, in which case the
prompt display disappears and the sample is run identified only by time and date.
Prompt sample ID? No
1-No 2-Yes
Press [ENTER] to confirm and return to Sample protocol setup.
Sampling protocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
Press [4] for Station# to assign the position at which the sample will be aspirated. The default
location is Station #2, the Test Tube Holder station. You may also elect to sample from Station
#3, the Manual Station or even Station #1, the Cal well. This last choice would be rarely used,
but does give you the flexibility to calibrate from a solution in the Test Tube Holder (#2) and
sample from the Cal Well (#1). If you have the YSI 2710 Turntable or YSI 2730 Monitor and
Control Accessory, stations #4 and #5 are also available. See the appropriate manual for
assignments.
Enter Sample Station#
> 2
Press [ENTER] to confirm and return to Sampling protocol setup.
Sampling protocol setup: 1-SipperHeight
2-Replicates 3-ID 4-Station# 5-Turntable
$)*# !
%+, ! !
Refer to the YSI 2710 Turntable Operation Manual for setup, operation and maintenance of the
turntable.
Press [0] to return to RUN mode setup.
RUN mode setup: 1-SampleProtocol
2-AutoStandby 3-AutoCal 4-Monitor