EasyManua.ls Logo

YSI 2700 SELECT - Page 82

YSI 2700 SELECT
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-13
Press [2] for AutoStandby to define the number of hours the unit will continue to update
calibration and be ready to sample. The default setting is 2. If you press [0] and confirm with
[ENTER], AutoStandby is disabled. The instrument then remains "sample ready" indefinitely.
You may enter any number between 0 and 999.
Enter autostandby time in hour
0 to disable> 2
Press [ENTER] to confirm and return to RUN mode setup.
RUN mode setup: 1-SampleProtocol
2-AutoStandby 3-AutoCal 4-Monitor
Press [3] for AutoCal to define parameters that initiate autocalibrations. The five categories are
shown in the display below. The default settings that initiate calibration are: 1) temperature drift
of more than 1°C since last cal, 2) after 15 minutes without cal update, 3) after 5 samples run
without a cal update, 4) after a probe current shift of 2% or greater compared to the last cal, and
5) after certain errors are detected (eg., baseline instability).
Autocal setup: 1-Temperature 2-Time
3-Sample 4-Cal shift 5-Sample error
You may alter any of these parameters to suit your application, however, you
may
compromise
precision and/or accuracy when doing so. YSI's stated specifications are based on the default
settings shown above.
-./0-'  ! $  "
 #  ! !1&
" &"$"$ 2  &&
$!&  #&$#" 
' $$$ 3 &  '
These selections are provided as part of the overall concept of the 2700 SELECT, flexibility.
You may have an application where 5% precision is acceptable if you can guarantee that the
instrument will not initiate a calibration at a critical sampling time. This menu allows you to
make that possible.
The 5 autocal parameter displays are shown below with their default values. The range of
values that may be entered is shown below each display. Press [ENTER] to confirm each
parameter and proceed to the next.
Press [1] for Temperature.
Enter allowable autocal temp drift in °C
0 to disable> 1
Confirm value shown or enter and confirm number of degrees from 1 to 20.

Table of Contents

Related product manuals