5-14
Autocal setup: 1-Temperature 2-Time
3-Sample 4-Cal shift 5-Sample error
Press [2] for Time.
Enter autocal time interval in minute
> 15
Confirm value shown or enter and confirm any number of minutes from 0 to 30000.
Autocal setup: 1-Temperature 2-Time
3-Sample 4-Cal shift 5-Sample error
Press [3] for Sample.
Enter number of samples between autocals
0 to disable> 5
Confirm value shown or enter and confirm any number of samples from 0 to 30000.
Autocal setup: 1-Temperature 2-Time
3-Sample 4-Cal shift 5-Sample error
Press [4] for Cal shift.
Enter allowable autocal shift in %
0 to disable> 2
Confirm value shown or enter and confirm any number from 0 to 100.
Autocal setup: 1-Temperature 2-Time
3-Sample 4-Cal shift 5-Sample error
Press [5] for Sample error.
Autocal on sample error? Yes
1-No 2-Yes
Press [ENTER] to confirm performing an Autocal whenever a sample error is detected, or press
[1] for No, then [ENTER] if you do NOT want the 2700 to perform an Autocal after detecting
an error.
Autocal setup: 1-Temperature 2-Time
3-Sample 4-Cal shift 5-Sample error