48
Zaxis Inc. LLC
801.264.1000
2442 South 2570 West
West Valley City, UT 84119
E
Event — The pressure change that signals the change in the device under tests condi-
tion. This trigger is used to end the test and compare the pressure reading to the limit
settings for pass/fail status.
F
Fail Light — The Isaac’s red indicator with an X-mark. The fail light turns on whenever
a test exceeds established parameters.
Firmware — The set of instructions stored in programmable read-only memory (Flash)
that controls Isaac’s operation. Firmware cannot be altered by the customer.
Firmware Version (How to nd) — The version of rmware running Isaac is dis-
played on the ‘About’ screen.
Fixturing — A xture is a device connected externally to the Isaac. Fixtures can be
mechanical, electrical, pneumatic, or combinations of all. Typical xtures are pneumatic
clamps that seal products during a pressure decay or ow test. The Isaac can supply
air from the coupling port to operate pneumatic xtures. Customers must specify at the
time of order whether they want coupling pressure to be line or test pressure. Many
xturing options are possible.
Flow — The amount of air passing through an object measured in cubic centimeters or
liters per time period (second, minute or hour) at a given pressure.
Flow Control — The Isaac has a built-in ow control to provide a slow pressure in-
crease (ramp up) needed for burst and crack testing. Users can precisely set the ow
control for the exact pressure build-up required for the product to be tested. After the
ow control is set for a particular product, further adjustment of the needle valve is
unnecessary.
Flow Standard — (1) A measuring instrument or certied restrictor that can be con-
nected to the Isaac as part of a ow calibration. The ow standard must have adequate
accuracy, stability, and repeatability needed to calibrate the Isaac. The ow standard
must have current calibration documentation if the customer requires accuracy trace-
ability. (2) A calibrated device to challenge the tester on an as needed basis. This de-
vice is calibrated and traceable. For example, a daily verication can be done to ensure
the tester will still nd the required ow value.