SERVICE
MANUAL
5.5
TEST
PROCEDURE
+
Set
an
oscilloscope
as
follows:
0.5
Volts
per
division.
0.2
Seconds
per
division.
+
Plug
in
the
mini-phone
plug
to
the
PD™-1400-series
device's
1
Valt
ECG
OUT
jack.
e
Connect
the
scope
probe
to
the
terminals
of
the
mini-phone
plug.
The
QRS
complex
display
should
appear
on
the
oscilloscope
having
an
amplitude
of
1
Volt
+/-
0.2
V
p-p.
Mark
the
Checkout
Procedure
Data
Sheet.
5.6
TEST
PROCEDURE
«
Set
ECG
SIZE
to
1x.
Depress
and
hold
the
ALARM
SET
UP
and
DOWN
arrow
buttons
to
activate
the
calibration
signal.
e
Press
the
RECORDER
ON/OFF
button.
The
strip
chart
should
display
a
signal
of
120
ppm
with
an
amplitude
of
10
mm
+/-1
mm.
M
Mark
the
Checkout
Procedure
Data
Sheet.
5.7
NOTCH
FILTER
TEST
PROCEDURE
e
With
the
PD™
1400-series
device
in
MONITOR
mode,
set
to
LEAD
I,
size
3x:
apply
a
1
mV
p-p
60Hz
sine
wave
from
the
simulator
to
the
ECG
input.
Note:
For
50Hz-configured
units,
input
a
50Hz
sine
wave.
Verify
that
the
waveform
on
the
recorder
is
less
than
1.5
mm.
Mark
the
Checkout
Procedure
Data
Sheet.
e
-
Turn
the
ECG
simulator
off.
1-12