PROCEDURES
7.
DEFIBRILLATOR
TESTS
EQUIPMENT
REQUIRED:
Dynatech
Nevada
Impulse
3000
or
equivalent
defibrillator
analyzer.
Stopwatch
WARNING
USE
EXTREME
CAUTION
IN
PERFORMING
THE
FOLLOWING
TESTS!!!
.
When
performing
this
check
using
paddles,
use
your
thumbs
to
operate
the
DISCHARGE
buttons.
No
portion
of
the
hand
should
be
near
the
paddle
plates.
Keep
hands
and
all
other
objects
clear
of
paddle
edges
when
discharging.
Limit
internal
discharges
to
100
Joules
or
less.
Do
not
discharge
the
PD™
1400-series
device
internally
more
than
3
times
in
one
minute.
Multiple
internal
discharges
at
more
than
100
Joules
may
damage
the unit.
All
persons
near
the
equipment
must
be
warned
to
“STAND
CLEAR”
prior
to
defibrillator
discharge.
74
TEST
PROCEDURE
Unit
should
be
off
Paddles
should
NOT
be
in
the
unit.
If
paddles
are
still
connected
to
unit,
disconnect
the
paddle
cable
from
the
unit.
Remove
paddles
from
paddle
well.
11-17