SERVICE
MANUAL
SECTION
2.
LEAKAGE
TEST
Perform
this
Action:
Look
for
Circle
the
Result
GND
Normal
GND
Lift:
22
ECG
Normal
-
—<10NA
|
YES
NO
Reversed
a
-
___
<i0HA
|
YES
NO
Pacer
Normal
=
_____<IQuA
|
YES
NO
_
Reversed
__
rr
<10uA
|
YES
NO
ECG
Risk
-
-
-<20uA
|YES
NO
Pacer
Risk
NO
Paddles
Normal
<100pA
|
YES
NO
Reversed
_
…-
_____<100uA
|
YES
NO
Paddle
Risk
se
70
<1000A
|
YES
NO
MEC
Normal
ーー
+
___<100uA
|
YES
NO
`
Reversed
__
<1004A
|
YES
NO
ECG
Risk
(unshielded)
NO
MEC
Risk
ーー
<1004A
|
YES
NO
SECTION.3.
DIAGNOSTICS
MODE
TESTS
Note:
These
tests
are
run
on
PD™1400
and
PD™2000
units
only.
Perform
this
Action:
Look
for:
Circle
the
Result
3.1
Power
up
the
unit
in
5
Beeps
heard.
YES
NO
diagnostics mode.
Battery
voltage
displayed.
YES
NO
3.2
Set
unit
to
PACER
ON.
Pace
rate
displays
30+/-8
|
YES
NO
Set
RATE
to
30,
80,
180
80+/-8
|
YES
NO
ppm.
180+/-8
|
YES
NO
33
Set
OUTPUT
to
20,80,
|
Pace
output
display
20+/-8
|
YES
NO
|
140
mA.
804/-8
|
YES
NO
140+/-8
|
YES
NO
1-28