1
3M
™
Attest
™
Super Rapid Readout
1
Biological Indicator 1492V
ProductDescription
The 3M™Attest™ Super Rapid
Readout Biological Indicator
1492V (brown cap, referred to
hereinafter as the 1492VBI)
is a self-contained biological
indicator specifically designed for
rapid and reliable qualification
testing and routine monitoring
of 270°F (132°C) and 275°F
(135°C) dynamic-air-removal
(pre-vacuum and steam-flush
pressure-pulse (SFPP)) steam
sterilization processes when
used in conjunction with a
3M™Attest™ Auto-reader
490(hereinafter referred to
as the 490Auto-reader) or a
3M™Attest™ Auto-reader 490H
having software version 4.0.0or
greater (hereinafter referred to
as the 490H Auto-reader). The
1492VBI is a single-use device.
A schematic illustrating the design of the 1492VBI is provided in
Figure 1. The self-contained design includes a carrier with spores of
Geobacillusstearothermophilus and a media ampoule containing bacteriological
growth medium which meets the requirements for growth promoting ability specified
in ISO 11138-1:2017. The spore carrier and media ampoule are contained in a
plastic vial topped with a brown cap. A chemical process indicator which changes
from pink to light brown or darker upon exposure to steam is located on the top
of the cap.
The 1492VBI utilizes the α-glucosidase enzyme system, which is generated
naturally within growing cells of Geobacillusstearothermophilus. The α-glucosidase
in its active state is detected by measuring the fluorescence produced by the
enzymatic hydrolysis of a non-fluorescent substrate, 4-methylumbelliferyl-α-D-gl
ucoside (MUG). The resultant fluorescent by-product, 4-methylumbelliferone (MU),
is detected in the Auto-reader. The presence of fluorescence within the specified
incubation time of the 1492VBI in the 490or 490H Auto-reader indicates a steam
sterilization process failure.
The 1492VBI can also indicate the presence of G.stearothermophilus organisms by
a visual pH color change reaction. Biochemical activity of the G.stearothermophilus
organism produces metabolic by-products that cause the media to change color
from purple to yellow which also indicates a steam sterilization process failure. Use
of this indication method is optional and is typically restricted to special studies.
1492VBIs meet ISO 11138-1:2017and ISO 11138-3:2017.
ReadoutTimes
The super rapid readout and the optional 48-hour visual pH color change incubation
times have been correlated with a 7-day incubation period following the FDA’s
Reduced Incubation Time protocol. Processed indicators were examined at 48hours
and 7days for detection of a visual pH color change. The fluorescent result and
the 48-hour visual pH color change readings were compared to the 7-day visual
pH color change readings to determine the readout time of the indicator. The time
to a fluorescent result is determined by the software version programmed on the
Auto-reader.
24-minute Fluorescent Result
1492VBIs incubated in a 490or 490H Auto-reader having software version
4.0.0or greater have a 24minute reduced incubation time result that correlates
to the 7day (168hours) visual readout result ≥ 97% of the time.
1-hour Fluorescent Result
1492VBIs incubated in a 490Auto-reader having a software version less than
4.0.0have a 1hour reduced incubation time result that correlates to the 7day
(168hours) visual readout result ≥97% of the time.
48-hour Visual pH Color Change Result
1492VBIs incubated in either a 490or 490H Auto-reader have 48hour reduced
incubation time results that correlate to the 7day (168hours) visual readout
result ≥97% of the time.
Due to the high reliability of the fluorescent result there is no advantage to
incubating 1492VBIs after the fluorescent result has been determined by the 490or
490H Auto-reader and recorded.
Note: To determine the software version on a 490or 490H Auto-reader, depress the
Information Button once. The Auto-reader will display two lines of text on the LCD
display. The top row indicates (left to right):
• Serial number
• Software version (V X.Y.Z)
• Current date and time.
PROCESS INDICATOR
BI CAP WITH LABEL
VIAL
MEDIA AMPOULE
AMPOULE CRUSHER
NONWOVEN MATERIAL
GROWTH CHAMBER
SPORE CARRIER
Figure 1: 3M™Attest™ Super Rapid Readout
Biological Indicator 1492V Design
5324306
5
02.21.18 PU file from 4614439, update GB to new Doc & insert FPO GB Copy for each language kmh
03.07.18 Collect InDesign files to be sent for translations kmh
05.04.18 Review document from translator, updated fonts for the Asian languages. Jer
05.14.18 Alts per marked up PDF kmh
05.17.18 Final release. TN