4.16 Operating Conditions
The medical laser NUVOLAS is not suitable for use in connection with combustible gas
mixtures of all kinds.
The device is not approved for operation at altitudes above 2,000 m above sea level. And
only for an air pressure between 1080 hPa and 750 hPa.
The following environmental conditions must be met:
• Ambient temperature: 18° C to 28° C
• Relative humidity: <75 % without condensation
4.17 Electromagnetic Compatibility
The NUVOLAS laser system meets the EMC requirements according to EN 60601-1-
2:2015. Guidelines and the manufacturer's declaration are described in Chapter 10.
A T T E N T I O N
Avoid using this device next to other devices or with other devices in a stacked form as
this could result in improper operation. If such use is still necessary, this device and the
other devices should be monitored to ensure that they are operating properly.