6.3 Programs and Settings
A T T E N T I O N
With some devices, programs and preset values are already stored. This does not
release the surgeon from checking the values according to his knowledge and the
desired interaction with the tissue. If necessary, the preset values must be adjusted.
6.3.1 Display / Main Menu
6.3.1.1 Operating mode
You can choose between pulse mode, single pulse mode (SP)
or continuous wave mode (CW) for the treatment. Just touch the
desired tab. Afterwards you can select the parameters as
needed.
6.3.1.2 Power (Watt)
The power controller is touch sensitive so that you can directly
choose the correct value. Alternatively, you can change the power
by using the arrow keys. The displayed power corresponds to the
power at the fiber tip with a tolerance of ±20 %.
6.3.1.3 Safe values
To safe changed values type on the floppy disc. The floppy disc is
only visible, if the parameters (power, pulse length, pulse pause) of
the application were changed. The following chapters explain in
more detail how to change the pulse length and pause.
Increase power
Power
Pulse counter
Decrease power
Standby / Ready
Button
Information
Back
Laser Log
Settings
Operating mode
Puls length
Pulse pause
Fleckgröße
Select user
Select application
Information settings
Aiming beam
brightness