8.3 Care by the user
The following care instructions can be carried out by the user. These serve to make your
work easier. For cleaning, the system must be disconnected from the mains. Always use
a damp, but never wet, soft cloth for cleaning and disinfecting.
For cleaning and disinfection, the power supply and the applicator must be disconnected.
In addition, the red dust protection cap should also be inserted into the fiber connector.
First clean the device with clear water and neutral detergent to remove coarse and visible
contamination. Make sure that no moisture penetrates into the interior.
Furthermore, wipe disinfection is possible. When choosing the disinfectant, pay attention
to the following:
• according to the manufacturer of the disinfectant, the disinfectant should be suitable
for non-invasive medical devices
• according to the manufacturer of the disinfectant, the disinfectant should be suitable
for wipe disinfection of surfaces
• the disinfectant should based on alcohol and/or quaternary compounds
• the disinfectant should be suitable for lacquers
• aldehyde-free disinfectant (recommended)
When wiping disinfection, the manufacturer's instructions for the disinfectant must be
observed. Below is a table of possible disinfectants that meet the above requirements: