EasyManuals Logo

A.R.C. Laser Nuvolas User Manual

A.R.C. Laser Nuvolas
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
User Manual NUVOLAS
April 2023
File: User Manual_NUVOLAS_rev3.docx 44
8.5.3 Error Codes
Following errors can be detected by the NUVOLAS during system check or use of the
device. If such an error occurs, restart the device. If the error still occurs contact A.R.C.
Laser service.
Error message
Short Description
Error message
Short Description
F00
Error system start
F23
Voltage test, error +12 V
F01
Power too low 20 %
F24
Voltage test, error +5 V
F02
Power too high 20 %
F25
Voltage test, error +24 V
F03
Laser radiation in
STAND-BY
F32
Error diode current
F04
Time out of opto coupler
F33
Peltier, temperature too high
F06
Short circuit foot switch,
left
F38
Error DA converter
F07
Short circuit foot switch,
right
F47
Error check sum
F11
Power setting is not
possibleh
F49
Error Chip combination
F15
Time out eye protection
filter
F50
IC²-Bus defect