6.8 Laser-Stop
If the Laser-Stop is pressed while starting the device, the devie won’t work, only the fan
starts. The display will remain black and no LEDs will light up.
If the laser stop is activated during treatment with NUVOLAS, the device is deactivated
immediately. The display and the LED will fade out.
Solution: Pull the Laser-Stop out. The power supply will be released againg and the
device can be started as usual, restarted respectively.
6.9 Switch Off
The NUVOLAS can be turned off by pressing the
on-/off-switch at the back of the device. You
should cover the display after each use, to
prevent the display from damage.
A T T E N T I O N
The Laser-Stop should only be used in emergencies.