2600T Series Pressure transmitters | OI/266/LowP-EN Rev.A 23
9 Error messages
9 Error messages
9.1 LCD Display
The LCD HMI in case of transmitter errors or malfunctioning is capable of displaying specific error/fault messages to help the user
in identifying the problem and resolve it. In case of an alarm, a message consisting of an icon and text appears at the bottom of
the process display. Use the (1) key to call up the information level. Use the “Diagnostics” menu to call up the error description
with a help text. In the error description, the error number is displayed in the second line (M028.018). Two further lines are used to
describe the error. The device status is divided into four groups. The message text beside this icon in the display provides
information about where to look for the error. There are the following areas: Electronic, Sensor, Configuration, Operating and
Process.
Icon Description
Error / Failure
Functional check (e.g. during simulation )
Out of Spec (e.g. operating with empty meter pipe)
Maintenance required
9.2 Error states and alarms
— Communication Board / Electronic related error messages.
Error message Tx LCD message Possible cause Suggested action Tx response
F116.023 Electronic Memory Failure Electronic memory corrupted The electronic must be replaced Analog Signal to Alarm
M026.024
NV Electronic Memory Burn
Error
Writings to the electronic non-
Volatile Memory has not been
successful
The communication board should be replaced
as soon as possible
no effect
—Sensor related error messages
Error message Tx LCD message Possible cause Suggested action Tx response
F120.016 Sensor Invalid
The sensor signal is not being updated correctly as a
result of an electronics failure, sensor error or a poorly
connected sensor cable.
Check cable connection,
check sensor and if problem
persists, the sensor must be
replaced.
Analog Signal to
Alarm
F118.017 Sensor Memory Fail Sensor memory corrupted The sensor must be replaced
Analog Signal to
Alarm
F114.000 P-dP Sensor Fail
Mechanical damage to the sensor. Loss of fill fluid from
the cell, ruptured diaphragm, broken sensor.
The sensor must be replaced
Analog Signal to
Alarm
F112.001
Static Pressure
Sensor Fail
The circuitry for the sampling of the static pressure has
failed.
The sensor must be replaced
Analog Signal to
Alarm
F110.002
Sensor Temperature
Fail
The circuitry for the sampling of the temperature has
failed.
The sensor must be replaced
Analog Signal to
Alarm
M028.018
NV Sensor Memory
Burn Error
Writings to the sensor non-Volatile Memory was not
successful
The sensor should be replaced
as soon as possible.
no effect
—Configuration related error messages.
Error message Tx LCD message Possible cause Suggested action Tx response
C088.030
Input Simulation
Active
The P-dP Value produced in output is
derived by the value simulated in input
Use a HART configurator (DTM - Hand held) to place
device back in to normal operating mode (Remove
the input simulation)
no effect
C088.030
Input Simulation
Active
The Static Pressure Value produced in
output is derived by the value simulated in
input
Use a HART configurator (DTM - Hand held) to place
device back in to normal operating mode (Remove
the input simulation)
no effect