PGC5000
PGC5000 Series Service Instructions 5 Diagnostics and troubleshooting
SI/PGC5000-EN, Rev B 24
INDICATOR ISSUE RESOLUTION (CHE CK & CORRECT) RESET SOURCE
DTC Di gital Input Electronic Se ction Purge Check purge pressure setting at AOC Ovn#DTC#.15
15 Alarm Input . o ve ns front panel gauge. Adjust as
necessary. Che ck i ns trum en t air
supply press ure. Che ck oven purge
pressure switch. Ensure all oven
doors are closed and secure.
DTC Di gital Input Purge Alarm Override Verify input device is functioning AOC Ovn#DTC#.16
16 properly.
Alarm Relay Open input overrides the Verify proper operation of common AOC Ovn#DTC#.D1
Override commo n alarm out put. Alarm override s witch.
Software Error Rol l up of sof twar e f aul ts Che ck t he EPC an d EPC Group Zi be AOC Ovn#.EPC
at the DTC more information.
Hardware Error Rol l up of de tec tor Che ck t he EPC an d EPC Group Zi be AOC Ovn#.EPC
diagnos ti cs at the EPC more information.
EP C Zon e (1-10) Rollup of Zone (1-10) Che ck E PC Zone spec ifi c indi ca tors AOC Ovn#.EPC(1-10)
Indic ator s and correct as nece ssary.
Prsr Sensor Fault Defective pressure sensor Replace pressure sensor if needed. AOC Ovn#.EPC.PZ.#
Lo w Pre sure Pre ssure out of range Adjust pressure zones low limit AOC Ovn#.EPC.PZ.#
(low) und er t he Set up Tab. Verify supply
pressure to the zone. Correct as
necessary. Replace EPC sensor or
solenoid valve as necessary.
Zones inlet f ilter plugged Replace a s necessary. Leak check
analyzer and supply lines.
High Pre ssure Pre ssure o ut of ran ge Adjust pressure zones high limit AOC Ovn#.EPC.PZ.#
(high) under t he Setu p Tab. Replace E PC
sensor or solenoid valve as
necessary.
Out of Control Pressure deviates >±3.575 Che ck anal ys is press ure contr ol TCF AOC Ovn#.EPC.PZ.#
psi from se tpoi nt settings. Che ck Zone i ndicat ors and
Diagnostic Dis plays.
Possible lea k Leak check analyzer
Ramp Out of Pressure deviates >±3.57 5 Che ck anal ys is press ure contr ol TCF AOC Ovn#.EPC.PZ.#
Contr ol psi from setpoint during settings. Che ck Zone i ndicat ors and
Pressure Ramp Diagnostic Dis plays.
Software Error Rollup of software faults at Che ck i ndicat ors and diag no sti c AOC Ovn#.DTM#
the De t Amp Displays at Detector A mplifier
Hardware Error Rollup of detector indicators Check detector indicators for more AOC Ovn#.DTM#
information and Diagnostic Displays
at Detector Amplifier.
D ete ct or Fa ul t TCD Master Digital board Detector amplifie r fault. Replace AOC Ovn#.DTM#.TCD.#
has detected a falut amplifier assembly.
Flame Out FID or FPD fla me not Check utilities, igniter including fuel, AOC Ovn#.DTM#.FID/FPD#
detected burner, c onne ction. Verify flows,
burner air.
Autoignite Limit Ret ry of Au toma tic Ignit es Check utilities, igniter including fuel, AOC Ovn#.DTM#.TCD.#
has been exceeded burner, c onne ction. Verify flows,
burner air.
Autzero Conflict Autozero attempted during Move autozero TCF to a time into AOC Ovn#.DTM#.TCD.#
peakmeasurement analysis when a component is not
being meas ured.
Offline (Confi gur ed) Strea m in offl ine s tate Inf orma ti on Only. AOC Strea m
Strea ms)