the screen during the session, the device is not RF‑compable. Once you have established telemetry,
select Interrogate.
The unit's Measured Data will be displayed on the FastPath™ Summary screen and should indicate normal
voltage and baery status, and the programmed parameters should be idencal to the Shipped Sengs
displayed on the programmer's on‑screen help.
Package Integrity. Ensure that the package has not been opened or in any way compromised. If damage is
suspected, return it to the manufacturer.
Use Before Date. Do not implant the device aer the Use Before Date printed on the label.
Opening the Package. If interrogaon of the device in its sterile packaging indicates normal funconing,
remove it from the package. The package's outer tray can be opened in non‑sterile surroundings. However,
when opening the inner tray, complete sterile technique must be observed.
Pre-Implant Tesng
Inducve Communicaon. Posion the Merlin
™ PCS or
Merlin
™ 2 PCS telemetry wand over the package and
select interrogate.
Pacing System Analyzer. Before implantaon, you may wish to test the device using a compable pacing
system analyzer (PSA) with calibrated sensivity and output sengs. When the probe is aached to the
device's connector, the programmed parameters should be idencal to the Shipped Sengs displayed on the
programmer's on-screen help.
Adaptor Probes. Use only IS‑1 PSA cable adaptor probes when tesng the device. Other probes may damage
the connector. Do not use IS‑1 adaptor probes in the IS4‑LLLL connector.
Compable Pacing Leads. Devices with IS‑1 connectors accept unipolar or bipolar IS‑1 short terminal pin
leads. Devices with IS4‑LLLL connectors accept IS4‑LLLL quadripolar leads. Prior to implantaon, make sure
leads t easily and snugly into the device header.
22