Guidance and manufacturer’s declaration –
electromagnetic immunity
TheSystemisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomeror
theuseroftheSystemshouldassurethatitisusedinsuchanenvironment.
IMMUNITY
test
IEC 60601
test level
Compliance
Level
Electromagnetic
environment – guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteor
ceramictile.Iffloorsarecovered
withsyntheticmaterial,therelative
humidityshouldbeatleast30%.
Electricalfast
transient/burst
IEC61000-4-4
±2kVfor
powersupply
lines
±1kVforinput/
outputlines
±2kVfor
powersupply
lines
±1kVforinput/
outputlines
Mainspowerqualityshouldbethat
ofatypicaldomestic,commercialor
hospitalenvironment.