IMMUNITY
test
IEC 60601
test level
Compliance
Level
Electromagnetic
environment – guidance
Surge
IEC61000-4-5
±1kV
differential
mode
±2kVcommon
mode
±1kV
differential
mode
±2kVcommon
mode
Mainspowerqualityshouldbethat
ofatypicaldomestic,commercialor
hospitalenvironment.
Voltage
dips,short
interruptionsand
voltagevariations
onpowersupply
inputlines
IEC61000-4-11
<5% U
T
(>95%dipin
U
T
)for0.5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dip
inU
T
)for5
seconds
<5% U
T
(>95%dipin
U
T
)for0.5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dip
inU
T
)for5
seconds
Mainspowerqualityshouldbethat
ofatypicaldomestic,commercial
orhospitalenvironment.Iftheuser
oftheSystemrequirescontinued
operationduringpowermains
interruptions,itisrecommended
thattheSystembepoweredfrom
anuninterruptiblepowersupplyor
abattery.
112