Blood Glucose Control
Solution Test
Control Solution Results
Compare the control solution result to the
range printed on the test strip instructions
for use. The result on your screen should be
in this range.
Control solution results are marked on
the results screen and in the Logbook
with a
symbol.
Note: Repeat the control solution test if the
results are outside of the range printed on
the test strip instructions for use. Stop using
the built-in meter if the control solution
results are repeatedly outside of the printed
range. Contact Customer Service.
Ketone Control Solution Test
5.6
mmol
L
22:23
Control Solution Test
0.4
mmol
L
22:23
Ketone Control
Solution Test