Using the ACCU-CHEK Aviva Plus System
Replace the code key and perform a control test every time you open a new box of test strips.
Refer to the test strip package insert for test strip storage and system operating conditions.
Store unused test strips in their original container with the cap tightly closed.
Close the container tightly immediately after removing a test strip. Moisture can damage the test strips
and produce incorrect results.
Do not remove test strips from the test strip container and put them into another container, such as a
plastic bag, pocket, purse, wallet, etc.
Discard the test strips if they are past the “Use By” date printed on the test strip container. If the “Use By”
date is missing or cannot be read, do not use the test strips. Contact the ACCU‑CHEK Customer Care
Service Center at 1‑800‑858‑8072.
Use the test strip immediately after removing it from the container.
Do not apply blood or control solution to the test strip before inserting it into the meter. If a result appears
before applying blood or control solution, do not act on that result.
Do not reuse test strips. Once blood or control solution has been applied to a test strip, discard it. If a
retest is necessary, use a new test strip.

Table of Contents

Other manuals for Accu-Chek Aviva

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva and is the answer not in the manual?

Accu-Chek Aviva Specifications

General IconGeneral
BrandAccu-Chek
ModelAviva
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals