Comparing Your Meter Result to a Laboratory Result
A common question is how the blood glucose results on your meter compare to the laboratory results. Your
blood glucose can change quickly, especially after eating, taking medication, or physical activity. If you test
yourself in the morning and then go to your healthcare professional’s oce for a blood glucose test, your
results will probably not match even if you are fasting. This is typically not a problem with your meter; it just
means that time has elapsed and your blood glucose has changed.
If you want to compare your meter result to the laboratory result, you must be fasting. Take your meter to
the healthcare professional’s oce, and test yourself by fingerstick within 5 minutes of having blood drawn
from your arm by a healthcare professional. Keep in mind that the laboratory uses dierent technology than
the meter, and that blood glucose meters for self‑testing generally read somewhat lower than the laboratory
result.
If you are fasting and you do a fingerstick test within 5 minutes of having your blood drawn, here are the
general guidelines to compare your meter result to the laboratory result:
If your blood glucose is below 75 mg/dL, your results generally should fall within ±15 mg/dL of the
laboratory result.
If your blood glucose is equal to or over 75 mg/dL, your results generally should fall within ±20 % of the
laboratory result.
50

Table of Contents

Other manuals for Accu-Chek Aviva

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva and is the answer not in the manual?

Accu-Chek Aviva Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
CodingNo coding required
Battery LifeApproximately 1000 tests
Memory500 test results
Battery Type1 x 3V lithium (CR2032)
Operating Humidity10% to 90% RH
Meter Weight59 g
Measuring Range10 mg/dL to 600 mg/dL
Automatic Shut OffAfter 90 seconds of inactivity