69
Cleaning and Disinfecting the Meter
To clean and disinfect without damaging the meter, follow these procedures carefully.
When to Clean and Disinfect the Meter
Clean the meter to remove visible dirt or other material prior to disinfecting.
Clean and disinfect the meter at least once per week and when blood is present on the surface of the
meter.
Clean and disinfect the meter before allowing anyone else to handle the meter. Do not allow anyone else
to use the meter on themselves for testing purposes.

Table of Contents

Other manuals for Accu-Chek Aviva

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva and is the answer not in the manual?

Accu-Chek Aviva Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
CodingNo coding required
Battery LifeApproximately 1000 tests
Memory500 test results
Battery Type1 x 3V lithium (CR2032)
Operating Humidity10% to 90% RH
Meter Weight59 g
Measuring Range10 mg/dL to 600 mg/dL
Automatic Shut OffAfter 90 seconds of inactivity