128
Acoustic Mode
6.5 Announcement of warnings and error messages
When the batteries will soon be exhausted or the last test strip in a
drum has been used, The meter issues an acoustic warning. This con-
sists of two short beeps sounded three times in succession
(
- - - - - - ).
The warning is given at different times.
Batteries are almost exhausted (the battery symbol is displayed,
see Chapter 9):
This acoustic warning is sounded when you turn the meter on.
You have used the last test strip in the drum (
End
and drum symbol
flash, see Chapters 2.3 and 12.2):
The acoustic warning is given after you have pressed the Test button
to turn the meter off.
If error messages (
E
with a number or the
EEE
, see Chapter 12.2) are
displayed, these are announced by two short beeps sounded four times
in succession (
- - - - - - - - ). This sequence is sounded only
once and is not repeated. The same sequence of beeps is sounded for
all error messages, i.e. they are not differentiated acoustically.
AC_CP_LCM_Manual_EN-UK final.indd 128_CP_LCM_Manual_EN-UK final.indd 128 20.6.2007 11:04:2820.6.2007 11:04:28