170
Symbols, error messages and troubleshooting
In the tables presented below you will first see the meanings of symbols
that appear singly or have the same meaning when they appear
together with other symbols. These are followed by the meanings of
combinations of symbols and of symbols that appear in the numeric
field.
12.1.1 Single symbols
Symbol Meaning
1
Draw up blood or control solution with the
test strip now.
2
During a test and when the result is being
displayed: the temperature is outside the
permitted range of +10 °C to +40 °C.
Do not use this result as a basis for thera-
peutic decisions.
Move to a place where the temperature is
between +10 °C and +40 °C and wait for
the temperature of the meter to adjust to
this temperature. Repeat the test with a
new test strip.
쑺쑺
AC_CP_LCM_Manual_EN-UK final.indd 170_CP_LCM_Manual_EN-UK final.indd 170 20.6.2007 11:04:4120.6.2007 11:04:41

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH

Related product manuals