EasyManuals Logo
Home>Accu-Chek>Blood Glucose Meter>Compact Plus

Accu-Chek Compact Plus Reference Manual

Accu-Chek Compact Plus
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #163 background imageLoading...
Page #163 background image
163
Measurement and storage conditions
11
11 Measurement and storage conditions
The reliability of your meter and finger pricker as well as the accuracy of
your results is influenced by a number of factors. Pay close attention to
the following points.
A meter that has been dropped can produce false results.
11.1 Temperature range
Without the batteries and without the test strip drum the meter
may be stored at between -25 °C and +70 °C.
With the batteries and without the test strip drum the meter may
be stored at between -10 °C and +50 °C.
At temperatures above +50 °C the batteries could leak and
damage the meter.
At temperatures below -10 °C the batteries do not have
enough power to keep the internal clock functioning.
With the batteries and with the test strip drum the meter may be
stored at between +2 °C and +30 °C.
For blood glucose tests and performance checks the permitted
range is from +10 °C to +40 °C.
AC_CP_LCM_Manual_EN-UK final.indd 163_CP_LCM_Manual_EN-UK final.indd 163 20.6.2007 11:04:4020.6.2007 11:04:40

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
BrandAccu-Chek
ModelCompact Plus
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals