178
Symbols, error messages and troubleshooting
Symbols Meaning
쑺쑺
앫
because you have just inserted a
partly used test strip drum,
앫 because you have just opened the
drum compartment cover.
Perform a performance check (see Chap-
ter 7). If the test produces a correct result,
you may continue to use the test strips. If
the result is incorrect, you need to insert a
new drum. Following the performance
check, the bottle symbol disappears from
the display. The drum symbol continues to
be displayed in order to draw your atten-
tion to the fact that the meter cannot cal-
culate the use-by period. The symbol only
disappears when you insert a new test
strip drum. All results are saved along with
the drum symbol. If you do not perform a
performance check, the flashing bottle
symbol appears with every test until you
insert a new drum.
AC_CP_LCM_Manual_EN-UK final.indd 178_CP_LCM_Manual_EN-UK final.indd 178 20.6.2007 11:04:4320.6.2007 11:04:43