15
Performing a Blood Glucose Test
Before performing a blood glucose test on patients, set up the meter correctly. The meter, a test strip, and a
disposable, single‑use lancet or blood collection device are required.
1. Wear protective gloves.
2. Wash or disinfect the patient’s nger and allow it to dry thoroughly. The blood drop does not spread so
easily over dry skin.
3. Use only a lancing device that is approved for use by healthcare professionals.
4. Use a lancing device with adjustable penetration depth settings. If using a Safe‑T‑Pro Plus Lancing Device,
select a penetration depth to suit the nature of the patient’s skin. Start with a shallow depth setting.
5. Select a puncture site along the side of a ngertip.
6. Insert a test strip into the meter.