Chapter 3: Control Testing for Healthcare Professionals
Introduction
Testing control solutions with known glucose levels establishes that the operator and the system are
performing acceptably. Control results must be within the dened acceptable ranges before valid patient
testing is allowed.
Testing Intervals
Follow your facility’s policy for control testing intervals.
Control testing should be performed:
The rst time before using the meter for patient
testing
At intervals established by the facility
When a new box of test strips is opened
If the test strip container is left open
If test strips were incorrectly stored
If there is a question about a patient’s glucose
result
To check the performance of the system
If the meter was dropped