22
When to Disinfect
Each time after you have tested a patient’s blood glucose, wipe the area around the test strip slot. If •
you nd blood on the cloth, wipe also the test strip slot itself and the display.
If there is blood on the meter, wipe the meter.•
If blood has entered the meter, wipe the area around the test strip slot and the display. Allow areas you •
have wiped time to dry thoroughly.
The meter, including the test strip slot and display must be thoroughly cleaned and disinfected at least •
once a day after use. Carefully check all recesses, grooves and gaps.
Suitable Disinfecting/Cleaning Solutions
Products suitable for use are:
70 % isopropyl alcohol•
Mild dishwashing liquid mixed with water•