Not enough blood or control solution
was drawn into the test strip for
measurement or was applied after
the test had started. Discard the test
strip and repeat the test.
The blood glucose may be extremely
high or a meter or test strip error
has occurred. If this conrms
symptoms, follow your facility
policy for treating questionable
blood glucose results. If it does not
conrm symptoms, repeat the test
and refer to Chapter 2, “Unusual Test
Results.” If this does not conrm symptoms, perform
a control test with the control solution and a new
test strip. If the control result is within the
acceptable range, review the proper testing
procedure and repeat the blood glucose test with a
new test strip. If the E‑3 code still appears for the
blood glucose test, the blood glucose result may be
extremely high and above the system’s reading
range. Follow facility policy. If the control result is
not within the acceptable range, see Chapter 3,
“Understanding Out‑of‑Range Control Test Results.”

Other manuals for Accu-Chek Performa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Performa and is the answer not in the manual?

Accu-Chek Performa Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Meter Dimensions94 x 52 x 21 mm
Measurement Range0.6-33.3 mmol/L
Memory Capacity500 results
Battery Type1 x CR2032 lithium coin cell
Automatic Power Off90 seconds after last user action
Operating Humidity10 to 90% RH
Storage Conditions10% to 90% RH
Meter Weight59 g (with battery)