52
Errors
Code Name Meaning Action
E6 Mechanical The pump has detected a • Press the Check key twice to turn
Error mechanical problem. off and confi rm the error.
• Remove or disconnect the infusion
Note: If this error continues, set from your infusion site.
contact Disetronic Pump • Remove the cartridge, adapter,
Support at 1-800-688-4578 infusion set (tubing), and battery
for more information. Contact from your insulin pump. Insert a
your healthcare provider for an new battery.
alternate therapy plan. • Rewind the piston rod.
• Install a new cartridge
and infusion set.
• Prime the infusion set.
• Reconnect the infusion set (tubing)
at your infusion site.
• Put you insulin pump into RUN mode.