Errors
Code Name Meaning Action
E7 Electronic The pump has detected an You cannot turn off and confi rm
Error electronic problem. this error by pressing the Check
key twice. You must remove the
battery instead.
Note: If this error continues, Remove or disconnect the infusion
contact Disetronic Pump set from your infusion site.
Support at 1-800-688-4578 Insert a new battery into the pump.
Error for more information. Contact Prime the infusion set.
your healthcare provider for Reconnect the infusion set (tubing)
an alternate therapy plan. at your infusion site.
Put you insulin pump into RUN mode.

Other manuals for Accu-Chek Spirit

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Spirit and is the answer not in the manual?

Accu-Chek Spirit Specifications

General IconGeneral
TypeInsulin Pump
ManufacturerRoche Diabetes Care
Insulin TypeRapid-acting insulin
Bolus TypesStandard bolus, Extended bolus, Multiwave bolus
Battery TypeAAA alkaline
Reservoir Capacity315 units
Bolus Dose Increments0.1 units
DisplayLCD
ConnectivityInfrared
Battery LifeApproximately 4-6 weeks