54
Errors
Code Name Meaning Action
E8 Power You either removed a battery Press the Check key twice to turn
Interrupt without putting your pump into off and confi rm the error.
STOP fi rst, or you dropped Your insulin pump goes into
your pump. STOP mode.
Check the time and date, and
Note: If this error occurs correct if needed.
during programming, check If a bolus and/or a Temporary Basal
your settings. Rate was interrupted by the error,
review the Bolus History and/or the
Temporary Basal Rate History for
the delivered amount and duration.
Put your insulin pump into RUN mode.
If necessary, program a new bolus
and/or Temporary Basal Rate.
E10 Cartridge The CHANGE THE CARTRIDGE Press the Check key twice to turn
Error function was carried out off and confi rm the error.
incorrectly. Remove or disconnect the infusion
set from your infusion site.
Move to the
CHANGE THE CARTRIDGE
menu.
Prime the infusion set.
Reconnect the infusion set (tubing)
at your infusion site.
Put you insulin pump into RUN mode.