55
Errors
Code Name Meaning Action
E11 Set Not You changed the cartridge and • Press the Check key twice to turn
Primed infusion set but did not prime off and confi rm the error.
the infusion set. • Remove or disconnect the infusion
set from your infusion site.
• Prime the infusion set.
• Reconnect the infusion set (tubing)
at your infusion site.
• Put your insulin pump into RUN
mode.
E12 Data Data transfer between your • Press the Check key twice to turn
Interrupted insulin pump and PC or off and confi rm the error.
Palm
®
device has been • Restart the data transfer.
interrupted.
E13 Language Language choice needs to • Press the Check key twice to turn
Error be checked. off and confi rm the error.
• Reset the language on your pump.
• Put your insulin pump into RUN mode.
Refer to ACCU-CHEK Spirit Insulin Pump User Guide for complete instruction, warnings, cautions and precautions.
CAUTIONS:
•
No insulin is delivered when an Error Code appears on the display. Immediate
action must be taken to correct the Error Code.
• Never prime infusion set that is connected to your body.