Errors
Code Name Meaning Action
E11 Set Not You changed the cartridge and Press the Check key twice to turn
Primed infusion set but did not prime off and confi rm the error.
the infusion set. Remove or disconnect the infusion
set from your infusion site.
Prime the infusion set.
Reconnect the infusion set (tubing)
at your infusion site.
Put your insulin pump into RUN
mode.
E12 Data Data transfer between your Press the Check key twice to turn
Interrupted insulin pump and PC or off and confi rm the error.
Palm
®
device has been Restart the data transfer.
interrupted.
E13 Language Language choice needs to Press the Check key twice to turn
Error be checked. off and confi rm the error.
Reset the language on your pump.
Put your insulin pump into RUN mode.
Refer to ACCU-CHEK Spirit Insulin Pump User Guide for complete instruction, warnings, cautions and precautions.
CAUTIONS:
No insulin is delivered when an Error Code appears on the display. Immediate
action must be taken to correct the Error Code.
• Never prime infusion set that is connected to your body.