3-10 Tags User Manual
Tag and Band Maintenance
Testing Tags
Accutech Tags operate by internal battery.
Over the course of normal operation, Tags
eventually lose battery power and the Tags
will need to be replaced. The Tag battery is
not replaceable. For maximum protection of
residents or assets, Accutech recommends
that Tags be tested on a weekly basis. There
are many ways that you can test Tags:
Enter a monitored zone
(Software will report.)
With a S-TAD
The Keypad’s Auxiliary LED
(Yellow) will light when a Tag is
detected (Optional; additional wire
required).
Check Visual Pulse LED if present
BR and CB Tags only - Remove
Band Alarm band or Cut Band band
(alarm will sound).
Cleaning Tags
All Tags are reusable but they must be
cleaned and sanitized between applications.
In long-term applications, periodically
replace the bands and clean the Tags. There
are a few methods that can be used to clean
and sanitize Tags:
HOWEVER, DO NOT:
DO NOT submerge Tags in water
DO NOT soak Tags in water
DO NOT submerge Tags in cleaning
solutions
DO NOT soak Tags in cleaning
solutions
DO NOT use solvents
DO NOT use abrasive cleaners or cloths
DO NOT put Tags in an AutoClave
DO NOT put Tags in a dishwasher
DO NOT steam clean Tags
DO NOT put Tags in a washing
machine or dryer
Acceptable Tag Cleaning Methods:
Seek guidance with your Infection
Control representative for approved
cleaners that can be used on
plastics
Wipe down with 3% hydrogen
peroxide with water solution then
rinse with a water-dampened cloth
towel and wipe dry. (wipe only,
DO NOT soak.
Antibacterial wipes then rinse with
a water-dampened cloth towel and
wipe dry (wipe only, DO NOT
soak.
Cleaning Bands
All Tag bands are for one-time use only. For
cleanliness and sanitary reasons, Accutech
recommends replacing applied bands on a
semi-monthly basis with a maximum of one
month between changes.
Storing Tags: To preserve battery life,
Tags must be turned off with an S-TAD
during storage or periods of non-use. BR
and CB Tags must be deactivated to prevent
alarms when not use. In addition, Tags
should be stored away from sources of
electrical noise (see example list below) and
stored in a metal container with lid.
Do not store Tags within 3 feet of any of the
following sources of electrical noise:
Computer monitors
Unshielded computer cables
Television sets
Medical Monitoring Equipment
X-ray and other imaging equipment
Fluorescent Lighting
Storing Bands
Extra Tag bands should be stored in clean
and dry environment.