11
Note: The Mission
®
Cholesterol Control Solution is intended for
validating cholesterol testing while using the Mission
®
Cholesterol Monitoring System. Both levels of control
solutions must be tested and fall within the assigned values
printed on the bottles.
Refer to the control solution package insert before using the controls. The
control solution bottle is labeled with the acceptable range that is specific for
that lot of control solution. The system is working properly if the control value
displayed by the meter is within the acceptable range printed on the bottle
label. If the value does not fall within the range, refer to the Control Solution
Package Insert for further instructions.
Precautions
Set the specimen type to blood (bL) before testing with the control solution.
Make sure the control solution and all the test materials reach operating
temperature of 20 - 40°C (68 - 104°F) prior to testing. The control solutions
and test materials are only accurate within this temperature range.
Use the control solution before the expiration date shown on the bottle.
Discard the control solution if it appears cloudy.
Use the Mission
®
Cholesterol Control Solution with the Mission
®
Cholesterol meter and test devices.
Make sure the control solution bottle is tightly closed before use.
The used device should be discarded according to local regulations after
testing.
Check the code chip before performing a test. Make sure to use the code
chip that is included with the box of test devices.
HDL/TRIG
Level 2
HDL/TRIG
Level 1
CHOL
Level 1
CHOL
Level 2