Test Device Precautions and Instructions for Use
Test Devices should be stored in their foil pouch to keep them in
working condition.
Do not store test devices outside of their package. Test devices must
be stored in the original package.
Do not transfer test devices to a new package or any other container.
For in vitro diagnostic use. Test devices are to be used only outside
the body for testing purposes.
Use the test device immediately after removing it from the foil pouch.
Do not use test devices that are torn, bent, or damaged in anyway.
Do not reuse test devices.
Before performing a test, make sure that the code number on the
meter display matches the number shown on the test device foil pouch
and on the ink-jet printing on the code chip.
Refer to the test device package insert for more details.
Control Devices
The Mission
®
Cholesterol CTRL Control Devices are gray and black devices,
which work with the Mission
®
Cholesterol Meter to ensure the optical system is
working properly. After the control device is inserted into the meter, the
meter’s optical system detects the color intensity of the reference pad in the
control device. The meter displays YES or no to indicate whether the meter
is functioning properly. Refer to Optical System Check for details.
The control device appears as shown below:
Precautions
Store in the closed canister at room temperature or in the refrigerator
within 2-30°C (36 – 86ºF). Avoid exposure to direct sunlight, extreme