6 Dispensing a treatment
6.1 Accessory usage conditions
The optical guide must be cleaned and sterilised prior to use. The rigid protection shield must be cleaned and
disinfected prior to use.
The MINILEDSUPERCHARGED accessories include:
l a charging base;
l an optical guide;
l a rigid protection shield.
Refer to the cleaning, disinfection and sterilisation protocols for accessories listed in the chapter Associated
documentation page 3.
6.2 Preparation for use
To prepare your MINILED, follow the steps below:
l Clean and disinfect the rigid protection shield with an alcohol wipe.
l Clean and disinfect the handpiece with an alcohol wipe.
l Sterilise the optical guide.
l Plug the charging base into the mains.
l Check that the translucent handpiece support flashes three times (red, green, blue) and that a beep sounds
when it is switched on.
l Remove the insulating ring located between the battery and the MINILED and screw the battery onto the
handpiece.
l Place the MINILED on its charging base.
l The MINILEDSUPERCHARGED is correctly installed when two beeps are heard and the translucent support
base lights up in blue and starts flashing.
l Leave the MINILED to fully charge.
l The MINILEDSUPERCHARGED is charged when the translucent support base stops flashing and remains lit up
in blue.
l Install the rigid protection shield.
l Wear safety goggles and protective gloves.
l Provide your patient with safety goggles.
User Manual|MINILED SUPERCHARGED|J02261EN|V2|(09)|08/2018|NG35EN010B - Page 13/38