10.5 Malfunction of the power tester
Possible causes Solutions
Defective light indicator Send the MINILED to Acteon's Customer Services team.
The optical guide is defective, damaged or dirty
Clean the optical guide using the multi-purpose syringe
air function and/or return the MINILED to Acteon's
Customer Service team
The reflector is defective or dirty
Clean the reflector using the multi-purpose syringe air
function and/or return the MINILED to Acteon's Customer
Service team
The power tester window is defective or dirty
Clean the power tester window and/or return the
MINILED to Acteon's Customer Service team
10.6 Other malfunctions
If the MINILED SUPERCHARGED is not working for any other reason, contact the Acteon Customer Services Team.
If you need to return your MINILED, please ensure the optical guide and battery are suitably packed to prevent any
impact damage during transportation.
Keep the original packaging of the medical device and use it to return the device for servicing, maintenance or repair.