8 Disinfection and sterilising
The instructions relating to cleaning, disinfection and sterilisation protocols for accessories supplied by SATELEC, a
company of Acteon group have been approved for each medical device and accessory. The applicable guides are listed
in the chapter Associated documentation page 3.
They can be downloaded at the following address: www.satelec.com/documents.
In all cases, the local regulations in force relating to the cleaning, disinfection and sterilisation protocols for
accessories take precedence over the information provided by SATELEC, a company of Acteon group.
8.1 Clean and disinfect the medical device
The MINILED must be OFF during cleaning and disinfecting procedures. It must also be disconnected from its electricity
supply.
Avoid using cleaning and disinfection products that contain flammable agents. Otherwise, ensure that the product has
completely evaporated or that there is no fuel left on the medical device and its accessories before switching it on.
Do not use an abrasive product to clean the medical device.
Never apply sprays directly to the medical device to clean it. Always spray the product onto a wipe, then clean
the medical device.
Use alcohol disinfectant wipes.
8.2 Cleaning, disinfecting and sterilising accessories
The MINILEDSUPERCHARGED accessories include:
l a charging base;
l an optical guide;
l a rigid protection shield.
The optical guide must be cleaned and sterilised prior to use. The rigid protection shield must be cleaned and
disinfected prior to use.
User Manual|MINILED SUPERCHARGED|J02261EN|V2|(09)|08/2018|NG35EN010B - Page 19/38