9 Monitoring and routine maintenance
The only preventive maintenance the medical device requires is:
l Monitoring of accessories
l Routine cleaning, disinfection and sterilisation
l Cleaning
Check the cleanliness of the handpiece nosepiece. It must be clean, smooth and corrosion-free. The optical guide must
fit easily and firmly inside it.
Before and after use, check the medical device and its accessories entirely for any problems. This is necessary to
detect any electrical isolation fault or damage. If necessary, replace damaged parts.
9.1 Check the power
It is important to regularly check that the lamp is working correctly.
This can be done using a purpose-designed tester.
To do this, proceed as follows:
1. Check that the optical guide is intact and has no composite residues.
2. Place the optical guide flat on the power tester.
3. Switch on the MINILED.
The power tester may yield the following results:
Colour Result
Green The lamp is working correctly
Red The lamp is not working correctly. Please read the chapter page 23
User Manual|MINILED SUPERCHARGED|J02261EN|V2|(09)|08/2018|NG35EN010B - Page 21/38