10 Identifying incorrect operation
In the event of incorrect operation, refer to the tables below to quickly identify and repair the non-complex parts of the
medical device.
If the incorrect operation is not described in the tables below, please contact your supplier or the customer service
team at SATELEC, a company of Acteon group.
Do not use the medical device if it appears to be damaged or faulty. Isolate the medical device and make sure that it
cannot be used.
The medical device will need to be sent away for repair.
10.1 Not working
Possible causes Solutions
The on/off button is set to off Press the on/off button to switch on the MINILED
The battery is flat Charge the battery
Symptoms: The MINILED does not switch on and does not emit blue light but the LCD screen is on.
Possible causes Solutions
The MINILED SUPERCHARGED is defective Send the MINILED to Acteon's Customer Services team.
The MINILED controls are blocked Disconnect and reconnect the battery.
10.2 Charging base not working
Possible causes Solutions
The wall socket is defective Contact your electrician.
The mains adapter is defective Send the MINILED to Acteon's Customer Services team.
The light indicators and/or beep sounds are defective Send the MINILED to Acteon's Customer Services team.
There is a contact fault at the jack connector Send the MINILED to Acteon's Customer Services team.
The fuse is defective Send the MINILED to Acteon's Customer Services team.
10.3 Optical Guide
Possible causes Solutions
Cure composite residue remains on the optical guide
The optical guide is damaged or is not clean
Clean the optical guide with an alcohol wipe.
Clean the optical guide using the multi-purpose syringe
air function.
Change the optical guide if necessary.
10.4 The power is not as expected
Symptoms:The composite does not cure.
Possible causes Solutions
The composite is too old or has been poorly preserved Use a new composite
The MINILED does not produce enough power
Check the power. Please read the chapter Check the power
page 21
The end of the optical guide is too far from the cure site
Place the end of the optical guide at 2 mm from the cure
site
User Manual|MINILED SUPERCHARGED|J02261EN|V2|(09)|08/2018|NG35EN010B - Page 23/38